Imaging & Therapeutic Technology |
1 From the Department of Radiology, David Geffen School of Medicine, University of California, Los Angeles, Box 951721, Los Angeles, CA 90095-1721. From the AAPM/RSNA Physics Tutorial at the 2001 RSNA scientific assembly. Received July 18, 2002; revision requested August 13 and received August 29; accepted August 30. Address correspondence to the author (e-mail: mmcnittgray@mednet.ucla.edu).
| Abstract |
|---|
|
|
|---|
© RSNA, 2002
Index Terms: Computed tomography (CT), radiation exposure Dosimetry Radiations, exposure to patients and personnel Radiations, measurement
| Introduction |
|---|
|
|
|---|
The purpose of this article is to review basic radiation dose concepts as well as those specifically developed to describe the radiation dose from CT. This involves describing the basic concepts of exposure, dose, and effective dose and showing how these are defined or estimated for CT. After the definitions are given, the effects of various technical CT factors on radiation dose are described. Next, some typical values for each dose parameter are given for a few select imaging protocols. Finally, some relevant resources and current activities are discussed.
| Radiation Dose Measures: General Definitions |
|---|
|
|
|---|
Absorbed Radiation Dose
Absorbed radiation dose (3), often referred to as radiation dose, describes the amount of energy absorbed per unit mass at a specific point. It is measured in grays (1 Gy = 1 J/kg) or rads (1 rad = 100 erg/g). The conversion between rads and grays is 100 rad = 1 Gy. Absorbed dose essentially describes how much energy from ionizing radiation has been absorbed in a small volume centered at a point; it does not describe where that radiation dose is absorbed or reflect the relative radiosensitivity or risk of detriment to those tissues being irradiated.
Effective Dose
Effective dose (35) (formerly referred to as the effective dose equivalent [6]) takes into account where the radiation dose is being absorbed (eg, which tissue has absorbed that radiation dose) and attempts to reflect the equivalent whole-body dose that results in a stochastic risk that is equivalent to the stochastic risk from the actual absorbed dose to those tissues irradiated in a nonuniform, partial-body irradiation such as a CT scan. It is a weighted average of organ doses, as described in Equation (1):
|
|
where E is the effective dose, wT is the tissue-weighting factor, wR is the radiation-weighting coefficient (1 for x rays), DT,R is the average absorbed dose to tissue T, T is the subscript for each radiosensitive tissue, and R is the subscript for each type of radiation (here, only x rays are present). The weighting factors are set for each radiosensitive organ in Publication 60 of the International Commission on Radiological Protection (ICRP) (4). Effective dose is measured in sieverts (Sv) or rems. The conversion between sieverts and rems is 100 rem = 1 Sv.
Although methods to calculate the effective dose have been established (ICRP Publications 26 [6] and 60 [4]), these methods depend heavily on the ability to estimate the dose to radiosensitive organs from the CT procedure (DT,R). However, determining the radiation dose to these organs is problematic and direct measurement is not possible.
| Radiation Dose Measures: CT Specific |
|---|
|
|
|---|
Variations within the Scan Plane
Projectional radiographic exposures are taken from one source position and the entrance skin dose is much larger than the exit skin dose, creating a large radiation dose gradient across the patient (Fig 1). In contrast, the tomographic exposure of CT scans with a full 360° rotation results in a radially symmetric radiation dose gradient within the patient. That is, in a uniform circular object, such as a test phantom, all of the points at a certain radius from the center have the same (or nearly the same) radiation dose (Fig 2). As we shall see, the magnitude of that dose gradient (the size of the difference from center to periphery) will be affected by several factors, including the size of the object, the x-ray beam spectrum, and the attenuation of the material or tissue.
|
|
|
|
|
To account for the effects from multiple scans, several dose descriptors were developed. One of the first was the Multiple Scan Average Dose (MSAD) descriptor (7,8,14). This is defined as the average dose resulting from a series of scans over an interval I in length:
|
|
where I is the interval of the scan length and Dseries(z) is the dose at position z parallel to the z (rotational) axis resulting from the series of CT scans.
Following this was the Computed Tomography Dose Index (CTDI) (15). This was defined as the radiation dose, normalized to beam width, measured from 14 contiguous sections:
|
|
where n is the number of sections per scan, T is the width of the interval equal to the selected section thickness, and Dsingle(z) is the dose at point z on any line parallel to the z (rotational) axis for a single axial scan. This index was suggested by the Food and Drug Administration and incorporated into the Code of Federal Regulations (16).
However, to be measured according to the definition, only 14 sections could be measured and one had to measure the radiation dose profiletypically done with thermoluminescent dosimeters (TLDs) or film, neither of which was very convenient. Measurements of exposure could be obtained with a pencil ionization chamber, but its fixed length of 100 mm meant that only 14 sections of 7-mm thickness could be measured with that chamber alone. To measure CTDI for thinner nominal sections, sometimes lead sleeves were used to cover the part of the chamber that exceeded 14 section widths.
To overcome the limitations of CTDI with 14 sections, another radiation dose indexCTDI100 was developed. This index relaxed the constraint on 14 sections and allowed calculation of the index for 100 mm along the length of an entire pencil ionization chamber (14), regardless of the nominal section width being used. This index is therefore defined as follows (9):
|
|
where N is the number of acquired sections per scan (also referred to as the number of data channels used during acquisition) and T is the nominal width of each acquired section (which is not necessarily the same as the nominal width of the reconstructed section width).
(Note: The product of N · T is meant to reflect the total nominal width of the x-ray beam during acquisition. Therefore, if a multidetector scanner is used with four channels [rows] of 1.25 mm each for scan acquisition, then, regardless of the reconstructed section width, NT = 4 · 1.25 mm = 5 mm; similarly, if the same scanner is used with four channels of 5-mm width for scan acquisition, then, regardless of the reconstructed section width, NT = 20 mm.)
Because the ionization chamber measures an integrated exposure along its 100-mm length, this is equivalent to the following formula:
|
|
where f is the conversion factor from exposure to a dose in air (use 0.87 rad/R), C is the calibration factor for the electrometer, E is the measured value of exposure in roentgens acquired from a single 360° rotation with a beam profile of NT (as defined earlier), L is the active length of the pencil ionization chamber, and N and T are as defined for Equation (4).
Thus, the exposure measurement, performed with one axial scan either in air or in one of the polymethyl methacrylate phantoms for which CTDI is defined, results in a calculated dose index, CTDI100. This index can be measured and calculated for the center location as well as at least one of the peripheral positions (1 cm below the surface) within the phantom to describe the variations within the scan plane as well.
CTDIw was created to represent a dose index that provides a weighted average of the center and peripheral contributions to dose within the scan plane (9). This index is used to overcome the limitations of CTDI100 and its dependency on position within the scan plane. The definition is as follows:
|
|
One final CTDI descriptor takes into account the parameters that are related to a specific imaging protocol, the helical pitch or axial scan spacing, and is defined as CTDIvol:
|
|
where N and T are as defined earlier and represent the total collimated width of the x-ray beam and I is the table travel per rotation for a helical scan or the spacing between acquisitions for axial scans.
For helical scans, the following formulas are used:
|
|
and
|
|
where pitch is defined as table distance traveled in one 360° rotation/total collimated width of the x-ray beam.
Another dose descriptor that is related to CTDI and is commonly reported on CT scanners and in the literature is the dose-length product (DLP) (9). This value is simply the CTDIvol multiplied by the length of the scan (in centimeters) and is given in units of milligray-centimeters:
|
|
This descriptor is used in one approach to obtain an estimate of effective dose that will be described later.
These CTDI descriptors are obviously meant to serve as an index of radiation dose due to CT scanning and are not meant to serve as an accurate estimate of the radiation dose incurred by an individual patient. Although the phantom measurements are meant to be reflective of an attenuation environment somewhat similar to a patient, the homogeneous polymethyl methacrylate phantom does not simulate the different tissue types and heterogeneities of a real patient. In addition, the CTDI100 calculation uses the f factor (from Eq [5]) only to convert from exposure to a dose in air; other tissues have different f factors. The f factor (in units of rads per roentgen) is determined by the ratio of the mass energy absorption coefficient of a tissue to that of air:
|
|
where µt/
t is the mass energy absorption coefficient of the tissue (eg, bone, lung, soft tissue) and µa/
a is the mass energy absorption coefficient of air. The mass energy absorption coefficient depends not only on the tissue but also on the energy of the photons, especially in the energy range used by CT. Thus, the CTDI100 calculation presents a very simplified condition for measuring radiation dose.
| Factors That Influence Radiation Dose from CT |
|---|
|
|
|---|
Beam Energy
The energy of the x-ray beam has a direct influence on patient radiation dose. This is selected by the operator (technologist) when the kilovolt peak is chosen for the scan. However, it is also influenced by the filtration selected for the scan. On some scanners, the selection of filtration is explicit; for others, it is implied (eg, by selection of the scan field of view [SFOV]). The influence of beam energy is shown in Table 1. When all other technical parameters are held constant and the kilovolt peak is increased on a single-detector CT scanner, the CTDIw values also increase for both the head and body CTDI phantoms. For example, when the kilovolt peak was increased from 120 to 140 on a CT/i scanner (GE Medical Systems, Milwaukee, Wis), the CTDIw increase was 37.5% for the head phantom and 39% for the body phantom.
|
|
Helical Pitch
For helical scans, the pitch parameter (defined as table distance traveled in one 360° rotation/total collimated width of the x-ray beam) has a direct influence on patient radiation dose. This is essentially because as pitch increases, the time that any one point in space spends in the x-ray beam is decreased. The relationship between radiation dose and pitch has been shown previously by using phantoms and thermoluminescent dosimeters (18). On the basis of these results, the CTDIvol (which is the only CTDI descriptor that takes pitch into account) varies as shown in Table 3, which gives the results when the pitch is varied and all other technical parameters are held constant on a single-detector CT scanner.
|
|
|
Effects of Object (and Patient) Size
In each of the preceding sections, we reported results for both the head and body phantoms. These phantoms, as described earlier, are made of the same soft-tissueequivalent material but are 16-cm-diameter and 32-cm-diameter right circular cylinders, respectively. To produce each table, we used the same technical factors for each phantom. Therefore, the primary difference in results between the head and body phantoms is size. Each of the tables shows that when the same technical parameters are used, the appropriate index shows that the smaller object always absorbs the higher dose and that the difference is at least a factor of two. Thus, for the same exposure factors, smaller patients would be expected to absorb much higher amounts of radiation dose than larger patients. This has significant implications for pediatric patients and small adults.
This is primarily because tissues are being exposed with both entrance radiation (as the tube is positioned directly over the tissue) and exit radiation (as the tube moves to the other side of the patient) as the source moves around the patient. For smaller patients, the exit radiation has been attenuated by less tissue and therefore is closer to the entrance radiation in its intensity, resulting in a much more uniform dose distribution (nearly equal at all locations in a 16-cm-diameter phantom). For the larger patient, the exit radiation is much less intense due to its attenuation through more tissue. This results in a difference within the scan plane with the higher radiation dose values occurring near the periphery, where entrance exposure is highest.
The effect of patient or object size on radiation dose has brought significant discussion into the proper selection of protocols for imaging pediatric patients as well as adjusting technical factors for patients according to size (2028).
Other Options for Reducing Scan Dose
In addition to the technical parameters discussed earlier, manufacturers have recently provided users with other means to reduce patient dose. One of these is an option to make changes in tube current based on the estimated attenuation of the patient at a specific location. Thus, the tube current will be programmed to a maximum value and can be reduced when there is information that a location along the patient is expected to be less attenuating than the most attenuating location to be imaged. This is determined by using both anteroposterior and lateral planning projection views. From these views, the tube current will be programmed to vary by location along the length of the patient and even as the tube is rotating around the patient. The exact details of the option vary by manufacturer.
In the near future, manufacturers may provide real-time (or close to real-time) tube current modulation (2931), so that tube current can be varied (reduced) as the scan is actually occurring, eliminating the need for both planning projections for dose reduction purposes (they may still be needed for planning purposes).
Indirect Effects
In addition to the direct effects that collimation has, as described earlier, there are some indirect effects that both it and the reconstruction algorithm may have on radiation dose. This is because, when thinner reconstructed image thicknesses are used, with all other factors held constant, there will be more noise in the image (where noise is defined as the standard deviation of the CT number). Therefore, noise typically increases with 1/
T, where T is the nominal section thickness. Therefore, a 10-mm-thick section is expected to have 3.2 times less noise than a 1-mm-thick section. Often when noisy images are obtained, the kilovolt peak or milliampere-seconds value or both are increased to offset the increase in noise due to narrower sections. Similar behavior is observed for the effects of the reconstruction algorithm. Algorithms that enhance higher spatial frequencies and improve spatial resolution (such as required for lung or skeletal imaging) also increase the noise in the image. To overcome this increase in noise, the kilovolt peak or milliampere-seconds value or both may be increased. This increase in kilovolt peak or milliampere-seconds value will result in an increase in radiation dose. Therefore, although changing the algorithm or section thickness may not have a direct effect on radiation dose, the selection of technical factors to offset the resulting increase in image noise may result in an increase in radiation dose.
Methods to Reduce Patient Dose
From the preceding discussion, it appears that there are several mechanisms to reduce the radiation dose to a patient (32,33). However, each of them has some resulting trade-off involved. Each of these is discussed below (27).
Reducing the Milliampere-Seconds Value.
From the results presented earlier, the radiation dose is linear with the milliampere-seconds value when all other factors are held constant. So, if the milliampere-seconds value is reduced by 50%, the radiation dose will be reduced by the same amount. However, this reduction will increase image noise by 1/
(mAs), which means that a 50% reduction in the milliampere-seconds value results in a noise increase of 41% (1/
2 = 1.41, a 41% increase). Depending on the requirements of the clinical application, this reduction may be readily accepted; in other cases, this type of reduction in milliampere-seconds may compromise the diagnostic quality of the imaging examination. For example, detection of high-contrast objects in the lung may not require a low-noise imaging protocol and the reduction in milliampere-seconds may be well tolerated. On the other hand, imaging low-contrast lesions in the liver does require a low-noise imaging protocol and the reduction in milliampere-seconds may limit the ability to detect these lesions.
Increasing Pitch. The radiation dose is inversely proportional to pitch when all other factors are held constant. Therefore, increasing pitch is one consistent way to reduce radiation dose. The trade-off in increasing pitch is an increase in effective section thickness, which results in increased volume averaging, which in turn may reduce the image signal (contrast between some object and background). The ability to use this type of dose reduction again depends on the clinical application.
Varying the Milliampere-Seconds Value by Patient Size. CT is an example of a digital modality in which the image quality continues to improve as the exposure increases. This is contrasted with analog projectional film, in which too high of an exposure results in an overexposed (too dark) film. Thus, when pediatric patients or small adult patients are imaged with CT using full-sized adult techniques, there is no penalty to image quality; in fact, the image quality is better under these conditions, as more photons reach the detector and image noise is reduced. However, the radiation dose to the smaller patient is potentially higher than is necessary to obtain a diagnostic image. Therefore, significant effort has recently been put into developing size- and weight-based imaging protocols to reduce radiation dose to pediatric patients and small adult patients, so that radiation dose can be reduced while still achieving sufficient diagnostic image quality (2125,27). This has typically been in the form of a reduced milliampere-seconds value for reduced patient size and has led to the development of suggested technique charts for pediatric patients.
Reducing Beam Energy. As discussed earlier, reducing the beam energy results in reduced radiation dose when all other factors are held constant. This will increase the image noise, and contrast changes will occur with a change in kilovolt peak, increasing with lower kilovolt peak for most tissue interfaces but decreasing or changing very little for others. From CTDI100 results, and not from the CTDIw results, we can observe that the radiation dose gradient is larger from periphery to center in the body phantom at lower kilovolt peak settings. This implies a greater relative skin dose for patients when lower kilovolt peak settings are used.
| Estimating Effective Dose from CT |
|---|
|
|
|---|
The first is based on Monte Carlo simulations performed several years ago (3437). This work, performed by members of the United Kingdoms National Radiological Protection Board (NRPB), used Monte Carlo methods to simulate CT scanning around a previously developed mathematical patient model (MIRD V [38]). This hermaphrodite patient model uses geometric shapes (cylinders, spheres, cones) to approximate the shape and location of all radiosensitive organs in the body; other approximations were used for the brain and lens of the eye. Physical measurements of CTDI in air on many scanners were collected as part of a national dosimetry program (36); these data were used to convert from the normalized Monte Carlo data to absolute dose values. The simulation also involved modeling scanner geometry, spectrum, section thickness, and other properties with the patient model in the beamaccounting for primary and scatter radiation. From these simulations, radiation doses for each organ in each imaged section were calculated and could be tallied to estimate the organ dose for each organ. These organ doses could then be combined with appropriate weighting factors to estimate effective dose (or effective dose equivalent from ICRP 26). This work was completed before helical and multidetector scanners were commonplace and did not explicitly model helical scanning or the effects of multidetector scanners.
This work formed the basis for several software programs that have taken the results of these simulations and put an interface on them to allow users to input some technical parameters and calculate an effective dose. These include programs such as CTDOSE (from John Le Heron at New Zealands National Radiation Laboratory [johnleh@nrl.moh.govt.nz]) and the Microsoft (Redmond, Wash) Excelbased product from the Imaging Performance Assessment of CT (ImPACT) scanners group in the United Kingdom (39). The latter software package matches the desired scanner to one of those used in the original study on the basis of CTDI in air as well as CTDI in phantom measurements.
In a related but independent effort, Zankl et al (40) from the Gesellschaft für Strahlen- und Umweltforschung (GSF) performed simulations on two different mathematical sex-specific phantoms, "Adam" and "Eva" (41), which were based on the ICRP data on Reference Man. Similar to the MIRD V phantom, the GSF mathematical phantoms were described by using combinations of spheres, ellipsoids, cones, and parts and combinations of these shapes. The Monte Carlo simulations were performed modeling photon transport through the simulated patient by using three different energy spectra. The section thickness was 1 cm, and radiation doses were calculated for single sections at positions varying contiguously from the bottom of the trunk to the top of the head. The result was mean organ conversion factors for each organ and per single CT section of 1 cm width centered at a given location in the phantom. The mean dose to a single organ resulting from a particular CT scan was estimated by summing up the contribution to the organ dose from each relevant section. The relative organ doses were converted to absolute organ doses by using the air kerma free-in-air on the axis of rotation. As in the National Radiological Protection Board data, these organ doses can be combined with appropriate weighting factors to estimate effective dose (or effective dose equivalent from ICRP 26).
The GSF work formed the basis for a software program with a graphical user interface (42). This package (WinDose; Scanditronix/Wellhofer, Schwarzenbruck, Germany, www.wellhofer.com) has taken the results of the simulations and put an interface on them to allow users to input some technical parameters and calculate an effective dose. The output is an estimate of the effective dose given the scanner characteristics, body region, scan range, and other parameters. This program also outputs the effective dose in millisieverts as well as in units of natural background radiation per year.
Another approach was that of estimating the energy imparted, developed by Atherton and Huda (4347). Energy imparted is the measure of the total ionizing energy deposited in the patient during the CT examination and here is symbolized by
. On the basis of the simulation data of Jones and Shrimpton (36,37), the energy imparted was calculated for a mathematical anthropomorphic phantom (38). For different CT systems, the ratio of effective dose equivalent (HE, defined in ICRP 26 [6]) to energy imparted (HE/
) and the ratio of effective dose (E, defined in ICRP 60 [4]) to energy imparted (E/
) were calculated and plotted as a function of the position along the anthropomorphic phantom. High peaks in this function showed the position of organs with a higher radiosensitivity like the gonads, breast, or thyroid. Therefore, in this approach, effective dose is estimated by first estimating the energy imparted to the body region being scanned and then multiplying by the ratio factor for that particular region.
One other method of note to estimate the effective dose involves conversion factors for a general anatomic region as described by the European Guidelines on Quality Criteria for Computed Tomography (9), which are based on the work of Jessen et al (48). In this approach, the CTDIvol and distance are used to estimate the DLP, which is then multiplied by a region-specific conversion factor to estimate the effective dose. These conversion factors range from 0.0023 mSv/mGy · cm for the head region to 0.017 mSv/mGy · cm for the chest region and 0.019 mSv/mGy · cm for the pelvis. This approach obviously does not take into account any patient-specific or even examination-specific factors but provides an easily estimated value of effective dose.
For a typical chest scan performed with a single-detector scanner at 120 kVp, 250 mAs, 57-mm collimation, and a pitch of 1, the CTDI100, center would be 10 mGy, the CTDI100, periphery would be 18 mGy, the CTDIw would be 15 mGy, and the CTDIvol would be 15 mGy. If a length of scan of 25 cm is assumed, the DLP would be 375 mGy · cm; when the conversion factor for the chest is used, one estimate of effective dose would be 6.4 mSv.
For a typical head scan performed with a single-detector scanner at 120 kVp, 300 mAs, 5-mm collimation, and a pitch of 1, the CTDI100, center would be 40 mGy, the CTDI100, periphery would be 40 mGy, the CTDIw would be 40 mGy, and the CTDIvol would be 40 mGy. If a length of scan of 17.5 cm is assumed, the DLP would be 700 mGy · cm; when the conversion factor for the head is used, one estimate of effective dose would be 1.6 mSv. Thus, the effective dose for the head scan is considerably less than that for the chest scan, even though the CTDI values for the head scan are much higher. This is because fewer of the radiosensitive organs are irradiated.
| Relevant Activities and Resources |
|---|
|
|
|---|
The American Association of Physicists in Medicine (AAPM) has convened a task group (Task Group 23: CT Dosimetry) under its Diagnostic X-ray Imaging Committee to address issues related to CT radiation dose. This task group is addressing issues relating to standardizing descriptors, measurement methods, and education activities related to these. This is in addition to the task group on reference values created under the AAPMs Radiation Protection Committee (Task Group 7: Reference Values for Diagnostic X-ray Examinations), which reports values for head CT scans. Both of these task groups operate under the AAPMs Science Council.
In addition, the CT accreditation program of the American College of Radiology (ACR) requires sites applying for accreditation to measure and calculate CTDI100, CTDIw, CTDIvol, DLP, and effective dose for pediatric body, adult head, and adult body techniques. Although this will not be a pass-fail criterion at first, the ACRs program will eventually use radiation dose as part of its pass-fail criteria. This program does provide acceptable value limits for each protocol.
The Center for Devices and Radiological Health (CDRH) of the Food and Drug Administration has made some public recommendations on CT scanning for pediatric and small adult patients, which they have published on their Web site (27). In addition, the agency has completed its Nationwide Evaluation of X-ray Trends (NEXT) for CT, which was undertaken in 2000 (1). Preliminary results were presented at the 2001 conference of the Radiological Society of North America. In this survey, CTDI values were reported from many different scanners placed at different institutions across the United States. From this survey, mean values as well as ranges of values are being obtained for different types of scans. Publications on the results of this survey are being prepared.
| Conclusions |
|---|
|
|
|---|
Although there are many dose descriptors and methods for measuring radiation dose and effective dose from CT, there are still areas for further research. Recent investigations include examining the effects of patient size on radiation dose (with specific implications for pediatric patients) and methods for estimating patient-specific radiation dose (49,50), rather than estimating based on phantoms or standard mathematical patients such as the MIRD V. In the near future, there may be methods to estimate the radiation dose, organ dose, and effective dose from arbitrary CT scanning protocols by using patient-specific models.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
D. Tack and P. A. Gevenois Efforts for Lowering Radiation Dose Delivered with CT: Raising Arms, or Is There More? Radiology, November 1, 2008; 249(2): 413 - 415. [Full Text] [PDF] |
||||
![]() |
E. Angel, C. V. Wellnitz, M. M. Goodsitt, N. Yaghmai, J. J. DeMarco, C. H. Cagnon, J. W. Sayre, D. D. Cody, D. M. Stevens, A. N. Primak, et al. Radiation Dose to the Fetus for Pregnant Patients Undergoing Multidetector CT Imaging: Monte Carlo Simulations Estimating Fetal Dose for a Range of Gestational Age and Patient Size Radiology, October 1, 2008; 249(1): 220 - 227. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Huda, K. M. Ogden, and M. R. Khorasani Converting Dose-Length Product to Effective Dose at CT Radiology, September 1, 2008; 248(3): 995 - 1003. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Bajpai, K. H. Lee, B. Kim, K. J. Kim, T. J. Kim, Y. H. Kim, and H. S. Kang Differences in Compression Artifacts on Thin- and Thick-Section Lung CT Images Am. J. Roentgenol., August 1, 2008; 191(2): W38 - W43. [Abstract] [Full Text] [PDF] |
||||