(Radiographics. 2001;21:S71-S80.)
© RSNA, 2001
Helping the Hepatic Surgeon |
Hepatic Imaging with Multidetector CT1
Hoon Ji, MD, PhD,
Jeffrey D. McTavish, MD,
Koenraad J. Mortele, MD,
Walter Wiesner, MD and
Pablo R. Ros, MD, MPH
1 From the Department of Radiology, Brigham and Womens Hospital,Harvard Medical School, 75 Francis St, Boston, MA 02115. Recipient of a Certificate of Merit award for an education exhibit at the 2000 RSNA scientific assembly. Received February 2, 2001; revision requested March 2 and received April 4; accepted April 11. Address correspondence to H.J. (e-mail: hji@partners.org).
 |
Abstract
|
|---|
Combining the advantages of a multirow detector array with a fast gantry rotation time, multidetector computed tomographic (CT) scanners can acquire sections at a faster rate than was previously possible. As a result, multidetector CT permits scanning during multiple specific phases of intravenous contrast enhancement and the acquisition of very thin sections over a large area, allowing the creation of multiplanar reconstructions with high z-axis resolution. The authors present an imaging strategy for the diagnosis and staging of hepatic pathologic conditions that emphasizes the role of multidetector CT. Users must master several scanning parameters to obtain the best image quality. For hepatic CT, it is practical to use relatively narrow collimation, increasing the pitch as needed to cover the entire liver. The choice of reconstruction interval is dependent on the problem for which the study is being performed. Water is recommended as an oral contrast agent for nonaxial reconstructions, since high-attenuation oral contrast agents might degrade them. Appropriate scanning delays for hepatic CT are dependent on the contrast-agent injection strategy used. A triple-pass technique, highlighting the arterial, parenchymal, and portal venous phases of enhancement, is recommended.
Index Terms: Computed tomography (CT), technology, 761.12119 Computed tomography (CT), thin-section, 761.12119 Liver, CT, 761.12114, 761.12119
 |
Introduction
|
|---|
Multidetector computed tomography (CT) represents an advance in CT technology that involves use of a multiple-row detector array instead of the traditional single-row detector array used in spiral CT. This innovation allows scanning four to eight times faster than scanning with spiral CT (1). Because of this faster scanning ability, multidetector CT permits many new scanning techniques that were not possible with single-row helical CT (2,3). Applying multidetector CT to hepatic imaging promises both improved clinical utility of CT and a challenge for radiologists.
In this review, we present an imaging strategy that emphasizes the role of multidetector CT in the diagnosis and staging of hepatic pathologic conditions.
 |
Parameters for Acquisition of Hepatic Images
|
|---|
The best-attainable image quality of any given CT system is determined by many system-inherent factors. The degree to which this potential is actually reached in clinical practice, however, is determined by a number of scanning parameters that the user can modify: (a) collimation, (b) pitch, (c) reconstruction interval, (d) contrast media, and (e) timing of scanning. Optimizing these parameters to achieve the best possible image quality for a specific multidetector CT examination requires a thorough understanding of their interrelationships.
Collimation
The practical considerations with regard to choice of collimation include spatial resolution, image noise, and length of coverage. Collimation should be tailored to the purpose of the study (ie, the specific structures being imaged). Spatial resolution may be increased by means of a decrease in the collimator width, but decreased collimator width also results in increased image noise and decreased length of coverage (Fig 1) (4).

View larger version (157K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 1a. Liver cyst. Multidetector CT images obtained with 1-mm collimation with 1.25-mm section thickness (a) and 2.5-mm collimation with 5-mm section thickness (b) show a tiny low-attenuation lesion in the liver (arrow in a). Narrow collimation produces higher image resolution but more noise.
|
|

View larger version (149K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 1b. Liver cyst. Multidetector CT images obtained with 1-mm collimation with 1.25-mm section thickness (a) and 2.5-mm collimation with 5-mm section thickness (b) show a tiny low-attenuation lesion in the liver (arrow in a). Narrow collimation produces higher image resolution but more noise.
|
|
The reduced coverage encountered with narrow collimation, however, can potentially be overcome with an increase in pitch. Thus, several factors must be balanced to achieve the best result for a particular imaging problem.
Pitch
In helical CT, pitch is defined as the table feed during a 360° rotation of the tube-detector apparatus divided by the collimator width. For multidetector CT, there are two distinct versions of pitch, depending on whether the entire beam collimation is considered or the "detector" collimation (ie, beam/4) is used. For example, if 10-mm beam collimation with a 10-mm table feed per rotation is used in a four-detector-row scanner (ie, "detector" collimation to four 2.5-mm channels), the pitch would be 1 if beam collimation is considered or 4 if detector collimation is considered. The selection of pitch for multidetector CT is affected by such factors as the required length of coverage, reconstructed section thickness, and image noise (5,6).
For a given duration of scanning, minimization of collimation (thus, an increase in pitch) will allow a narrow reconstructed section thickness and potentially improve spatial resolution, particularly in nonaxial planes. However, maximization of pitch may also result in a decrease in contrast resolution. Thus, the most appropriate choice of scanning parameters is dependent on the imaging problem under consideration. For example, because CT angiography allows imaging of structures with very high attenuation against a background of much lower attenuation (ie, high inherent contrast), the loss of contrast resolution caused by maximization of pitch can be disregarded. For CT of the liver, however, the potential loss of contrast resolution with maximal pitch must be taken into consideration. For hepatic CT, because both good spatial resolution and good contrast resolution are required, a practical approach is to use relatively narrow collimation, increasing the pitch as needed to cover the entire liver.
Reconstruction Interval
The choice of reconstruction interval is dependent on the clinical problem for which the study is being performed. The smaller the reconstruction interval, the greater the longitudinal (z-axis) resolution, with a resultant loss in z-axis coverage (7). If it is expected that multiplanar reconstructions will be required, a small reconstruction interval with overlapping sections is advantageous (Fig 2). Another consideration is the detection of small lesions, as overlapping reconstructions have been shown to improve the detection of hepatic metastases. In most cases, if nonaxial reconstructions are not anticipated, contiguous reconstructions usually are adequate.

View larger version (151K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 2a. Hepatocellular carcinoma. (a) Coronal reconstruction image, generated from 5-mm axial sections with a 5-mm reconstruction interval, shows a heterogeneous-attenuation mass in the tip of the right lobe with portal vein thrombosis (arrowhead). (b) Sagittal multiplanar reconstruction image shows rippling artifact on the liver surface as well as destruction of a lumbar spinal body by metastasis (arrow).
|
|

View larger version (148K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 2b. Hepatocellular carcinoma. (a) Coronal reconstruction image, generated from 5-mm axial sections with a 5-mm reconstruction interval, shows a heterogeneous-attenuation mass in the tip of the right lobe with portal vein thrombosis (arrowhead). (b) Sagittal multiplanar reconstruction image shows rippling artifact on the liver surface as well as destruction of a lumbar spinal body by metastasis (arrow).
|
|
A practical trade-off for improved resolution is slower processing of data, the need for more data storage, and increased time for interpretation of studies (8). These issues have become less problematic with improvements in computing speed and the increasing use of soft-copy interpretation.
Contrast Media
Several studies have shown that the volume of contrast medium used to enhance the thoracic vascular structures during helical CT can be as little as one-half that used for dynamic incremental CT (9,10). For abdominal CT, however, the issue of reduction in contrast medium has not been resolved. This issue must be revisited with multidetector CT because of its further increase in scanning speed. For liver imaging, the most important patient-related factor that affects the magnitude of hepatic contrast enhancement is body weight (11). The use of oral contrast medium may change with multidetector CT because of the increased use of nonaxial reconstructions and the potential for high-attenuation oral contrast medium to degrade the reconstructions. In these situations, water is recommended as an oral contrast agent.
Timing of Scanning
The appropriate scanning delays for hepatic CT are dependent on the strategy used for injecting the contrast medium. The timing of peak aortic and hepatic contrast enhancement is primarily dependent on the rate of injection (12,13). Rapid or low-volume (shorter-duration) injections produce earlier peak enhancement, whereas slow or high-volume (longer-duration) injections result in later peak enhancement. These factors must be taken into account before a fixed scanning delay is instituted.
A preliminary minibolus (5 mL/sec for 4 seconds), with scanning every 2 seconds beginning 10 seconds after the injection is started, can be used. The time to peak aortic enhancement is determined from the resultant time attenuation curve and is used to calculate the scanning delay. The scanning delay can also be accurately timed with a bolus tracking software program (14).
With the triple-pass technique, multidetector CT can be used to image during three distinct hepatic circulatory phases (1517). The first pass provides an image of the arterial phase, which exquisitely displays the hepatic arterial system, though it may be too early to see some hypervascular neoplasms (Fig 3). The second pass corresponds in timing to initial opacification of the portal venous system and is labeled the parenchymal or portal vein inflow phase. The enhancement of hypervascular neoplasms is maximized during this phase. In both primary and metastatic hypervascular neoplasms (eg, hepatocellular carcinoma, islet cell tumor, carcinoid, sarcoma), approximately 30% more lesions are detectable during this phase than during the later portal venous phase (Fig 4) (1721). During the third imaging pass, conventionally labeled the portal venous phase but also known as the hepatic venous phase, the hepatic veins are enhanced, having been unenhanced during both the early arterial and parenchymal phases. During this phase, enhancement of background hepatic parenchyma is maximized. Tumors that are hyperattenuating on the arterial-phase and parenchymal phase images may appear iso- or hypoattenuating on the hepatic venous-phase image (Fig 5) (1619).

View larger version (138K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 3a. Hepatic hemangioma. Triphasic contrast enhancement with multidetector CT shows sequential enhancement of hepatic vessels and tumor tissue. (a) Axial image obtained during the arterial phase of enhancement shows early eccentric enhancement of the focal hepatic lesion. Portal veins are not enhanced. (b) Image obtained during the parenchymal phase shows further enhancement of the lesion. Portal veins are enhanced, but hepatic veins are not. (c) Image obtained during the hepatic venous phase shows enhancement of hepatic veins and maximal enhancement of the hepatic parenchyma. This phase of enhancement is also known as the portal venous phase.
|
|

View larger version (148K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 3b. Hepatic hemangioma. Triphasic contrast enhancement with multidetector CT shows sequential enhancement of hepatic vessels and tumor tissue. (a) Axial image obtained during the arterial phase of enhancement shows early eccentric enhancement of the focal hepatic lesion. Portal veins are not enhanced. (b) Image obtained during the parenchymal phase shows further enhancement of the lesion. Portal veins are enhanced, but hepatic veins are not. (c) Image obtained during the hepatic venous phase shows enhancement of hepatic veins and maximal enhancement of the hepatic parenchyma. This phase of enhancement is also known as the portal venous phase.
|
|

View larger version (149K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 3c. Hepatic hemangioma. Triphasic contrast enhancement with multidetector CT shows sequential enhancement of hepatic vessels and tumor tissue. (a) Axial image obtained during the arterial phase of enhancement shows early eccentric enhancement of the focal hepatic lesion. Portal veins are not enhanced. (b) Image obtained during the parenchymal phase shows further enhancement of the lesion. Portal veins are enhanced, but hepatic veins are not. (c) Image obtained during the hepatic venous phase shows enhancement of hepatic veins and maximal enhancement of the hepatic parenchyma. This phase of enhancement is also known as the portal venous phase.
|
|

View larger version (154K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 4a. Multiple metastases in a 59-year-old woman. (a) Arterial-phase image shows indistinct, low-attenuation, atypical metastatic carcinoid tumors. (b) Image obtained during the late parenchymal phase allows clearer identification of multifocal, low-attenuation masses in both hepatic lobes. (c) Image obtained during the hepatic venous phase shows maximal parenchymal enhancement. Demarcation of the tumors is less distinct than that on the parenchymal-phase image.
|
|

View larger version (160K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 4b. Multiple metastases in a 59-year-old woman. (a) Arterial-phase image shows indistinct, low-attenuation, atypical metastatic carcinoid tumors. (b) Image obtained during the late parenchymal phase allows clearer identification of multifocal, low-attenuation masses in both hepatic lobes. (c) Image obtained during the hepatic venous phase shows maximal parenchymal enhancement. Demarcation of the tumors is less distinct than that on the parenchymal-phase image.
|
|

View larger version (157K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 4c. Multiple metastases in a 59-year-old woman. (a) Arterial-phase image shows indistinct, low-attenuation, atypical metastatic carcinoid tumors. (b) Image obtained during the late parenchymal phase allows clearer identification of multifocal, low-attenuation masses in both hepatic lobes. (c) Image obtained during the hepatic venous phase shows maximal parenchymal enhancement. Demarcation of the tumors is less distinct than that on the parenchymal-phase image.
|
|

View larger version (140K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 5a. Small hepatocellular carcinoma. (a) Arterial-phase image shows an enhancing mass in the liver (arrow). (b) Image obtained during the parenchymal phase clearly shows the mass. (c) On the image obtained during the hepatic venous phase, however, no definable mass is shown because the "temporal window" has passed. Hepatic scanning during three distinctive phases of enhancement may provide information useful for characterization of lesions.
|
|

View larger version (142K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 5b. Small hepatocellular carcinoma. (a) Arterial-phase image shows an enhancing mass in the liver (arrow). (b) Image obtained during the parenchymal phase clearly shows the mass. (c) On the image obtained during the hepatic venous phase, however, no definable mass is shown because the "temporal window" has passed. Hepatic scanning during three distinctive phases of enhancement may provide information useful for characterization of lesions.
|
|

View larger version (142K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 5c. Small hepatocellular carcinoma. (a) Arterial-phase image shows an enhancing mass in the liver (arrow). (b) Image obtained during the parenchymal phase clearly shows the mass. (c) On the image obtained during the hepatic venous phase, however, no definable mass is shown because the "temporal window" has passed. Hepatic scanning during three distinctive phases of enhancement may provide information useful for characterization of lesions.
|
|
Though all three phases are often useful in the initial detection and characterization of liver lesions, CT arteriography is usually not necessary in patients who are undergoing posttherapeutic care, and only the second and third imaging passes may be performed.
 |
Sample Protocol for Four-Detector-Row CT of the Liver
|
|---|
For multipass hepatic imaging in our department, 30% nonionic contrast material (30 g iodine) is administered at a rate of 4 mL/sec for 25 seconds. Three passes are performed, one each during the arterial phase (20 seconds after injection of contrast medium), parenchymal phase (35 seconds after the initiation of injection), and hepatic venous phase (60 seconds after the start of injection). Collimation is 2.5 mm. The thickness of reconstructed sections is 5 mm in all cases, with a 60% overlap used during the arterial phase. In all cases, a table speed of 15 mm/sec is used (Table).
 |
Clinical Utility of Multidetector CT
|
|---|
Improved Detection of Lesions
The shorter imaging times of multidetector CT allow imaging during periods of more intense enhancement, as well as more precise imaging during specific phases of enhancement (Figs 6, 7). In addition, with multidetector CT there may be less respiratory misregistration and greater z-axis resolution. The speed of coverage of four-detector-row CT is up to four times faster than that with a single-detector-row scanner. Thus, CT of the liver can be performed routinely with very thin collimation, yielding greater conspicuity of small lesions for better detection of lesions (Fig 1). Use of 2.5-mm-thick sections has been shown to result in a 46% increase in lesion detection compared with the use of 10-mm-thick sections and an 18% increase in detection compared with the use of 5-mm-thick sections (22).

View larger version (146K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 6a. Metastatic melanoma. (a) Multidetector CT scan obtained before administration of contrast medium shows a large, low-attenuation mass in the liver. (b) Scan obtained during the late arterial phase shows partial enhancement of the mass. (c) Scan obtained during the hepatic venous phase shows the mass with homogeneously low attenuation.
|
|

View larger version (159K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 6b. Metastatic melanoma. (a) Multidetector CT scan obtained before administration of contrast medium shows a large, low-attenuation mass in the liver. (b) Scan obtained during the late arterial phase shows partial enhancement of the mass. (c) Scan obtained during the hepatic venous phase shows the mass with homogeneously low attenuation.
|
|

View larger version (139K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 6c. Metastatic melanoma. (a) Multidetector CT scan obtained before administration of contrast medium shows a large, low-attenuation mass in the liver. (b) Scan obtained during the late arterial phase shows partial enhancement of the mass. (c) Scan obtained during the hepatic venous phase shows the mass with homogeneously low attenuation.
|
|

View larger version (129K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 7a. Hepatic perfusion abnormality in a 44-year-old man. (a) Multidetector CT scan obtained during the arterial phase shows transient focal enhancement (arrow). (b, c) Scans obtained sequentially during the parenchymal (b) and hepatic venous (c) phases show no mass. This case shows unusual masslike nodular enhancement. Localized, transient enhancement is common with triphasic scanning. The most common underlying causes are arterial-portal fistulas and minor portal venous variants.
|
|

View larger version (133K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 7b. Hepatic perfusion abnormality in a 44-year-old man. (a) Multidetector CT scan obtained during the arterial phase shows transient focal enhancement (arrow). (b, c) Scans obtained sequentially during the parenchymal (b) and hepatic venous (c) phases show no mass. This case shows unusual masslike nodular enhancement. Localized, transient enhancement is common with triphasic scanning. The most common underlying causes are arterial-portal fistulas and minor portal venous variants.
|
|

View larger version (128K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 7c. Hepatic perfusion abnormality in a 44-year-old man. (a) Multidetector CT scan obtained during the arterial phase shows transient focal enhancement (arrow). (b, c) Scans obtained sequentially during the parenchymal (b) and hepatic venous (c) phases show no mass. This case shows unusual masslike nodular enhancement. Localized, transient enhancement is common with triphasic scanning. The most common underlying causes are arterial-portal fistulas and minor portal venous variants.
|
|
Three-dimensional Imaging
With the rapid scanning ability of multidetector CT, it is feasible to obtain a three-dimensional data set of the entire liver during a single breath hold. With reconstruction of these data, high-quality three-dimensional images may be obtained. We have found that image quality in reconstructed sagittal, coronal, or curved planes is excellent in most cases. The use of curved multiplanar reconstruction along a dilated biliary tree or along vessels may more clearly elucidate the anatomic and pathologic characteristics than the reading of axial images alone. Such unique imaging approaches may ultimately improve lesion detection, characterization, and surgical planning (Fig 8) (23,24).

View larger version (166K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 8a. Liver at three-dimensional imaging. (a) Three-dimensional volume-rendered image shows a hemangioma in the right lobe of the liver and the adjacent vasculature. (b) Volumetric reconstruction image from CT angiography displays the hepatic and abdominal vessels and their interrelation. A large eccentric aortic aneurysm is seen near the origin of the right renal artery.
|
|

View larger version (173K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 8b. Liver at three-dimensional imaging. (a) Three-dimensional volume-rendered image shows a hemangioma in the right lobe of the liver and the adjacent vasculature. (b) Volumetric reconstruction image from CT angiography displays the hepatic and abdominal vessels and their interrelation. A large eccentric aortic aneurysm is seen near the origin of the right renal artery.
|
|
In addition, multidetector CT permits dynamic subtraction CT of the liver during a single breath hold. Not only is this method an elegant means to quantitate vascularity, it may correlate with the biologic activity of tumors (25).
 |
Limitations of Multidetector CT
|
|---|
The practical result of multidetector CT is the generation of many more narrowly collimated sections than are routinely obtained with single-detector-row CT. For certain examinations such as CT angiography, the image files may be extraordinarily large, with more than 1,000 images. As this number of images is impractical to view on film, soft-copy viewing at a workstation is the most efficient way to review these large data sets (8).
Timing of scanning becomes more critical as the increased speed of multidetector CT narrows the "temporal window" for desirable phase of enhancement. Without proper timing for abdominal studies, it is possible to scan too early and miss the intended phase. The distinct hepatic circulatory phases can be successfully imaged only through adjustment of the acquisition time to the circulation time of the individual patient.
Another potential drawback of multidetector CT may be an increased radiation dose, due to both the high milliampere-second levels used in many thin-section acquisitions to maintain image quality and the increased prevalence of multipass scanning. It behooves radiologists to make an effort to minimize radiation dose when possible. Although cone-beam artifacts are theoretically a potential problem of multidetector CT, they have not caused significant image degradation to date.
 |
Conclusions
|
|---|
Multidetector CT allows highly precise imaging during three (or more) distinct phases of hepatic enhancement. Optimal acquisition timing, in combination with thinner collimation, permits improved lesion detection and will also probably improve characterization of lesions. With the advances in rapid volume rendering and other three-dimensional techniques, a new era of CT-based three-dimensional imaging of the abdominal viscera is becoming a reality.
 |
References
|
|---|
-
Berland LL, Smith JK. Multidetector array CT: once again, technology creates new opportunities (editorial). Radiology 1998; 209:327-329.[Free Full Text]
-
Rydberg J, Buckwalter KA, Caldemeyer KS, et al. Multisection CT: scanning techniques and clinical applications. RadioGraphics 2000; 20:1787-1806.[Abstract/Free Full Text]
-
Klingenbeck-Regn K, Schaller S, Flohr T, Ohnesorge B, Kopp AF, Baum U. Subsecond multi-slice computed tomography: basics and applications. Eur J Radiol 1999; 31:110-124.[Medline]
-
Hu H, He HD, Foley WD, Fox SH. Four multidetector-row helical CT: image quality and volume coverage speed. Radiology 2000; 215:55-62.[Abstract/Free Full Text]
-
McCollough CH, Zink FE. Performance evaluation of a multi-slice CT system. Med Phys 1999; 26:2223-2230.[Medline]
-
Wang G, Vannier MW. The effect of pitch in multislice spiral/helical CT. Med Phys 1999; 26:2648-2653.[Medline]
-
Hu H. Multislice helical CT: scan and reconstruction. Med Phys 1999; 26:5-18.[Medline]
-
Rubin GD. Data explosion: the challenge of multidetector-row CT. Eur J Radiol 2000; 36:74-80.[Medline]
-
Costello P, Dupuy ED, Ecker CP, Tello R. Spiral CT of the thorax with reduced volume of contrast material: a comparative study. Radiology 1992; 183:663-666.[Abstract/Free Full Text]
-
Costello P, Ecker CP, Tello R, Hartnell GG. Assessment of the thoracic aorta by spiral CT. AJR Am J Roentgenol 1992; 158:1127-1130.[Abstract/Free Full Text]
-
Heiken JP, Brink JA, McClennan BL, Sagel SS, Crowe TM, Gaines MV. Dynamic incremental CT: effect of volume and concentration of contrast material and patient weight on hepatic enhancement. Radiology 1995; 195:353-357.[Abstract/Free Full Text]
-
Heiken JP, Brink JA, McClennan BL, Sagel SS, Forman HP, DiCroce J. Dynamic contrast-enhanced CT of the liver: comparison of contrast medium injection rates and uniphasic and biphasic injection protocols. Radiology 1993; 187:327-331.[Abstract/Free Full Text]
-
Chambers TP, Baron RL, Lush RM. Hepatic CT enhancement. II. Alterations in contrast material volume and rate of injection within the same patients. Radiology 1994; 193:518-522.[Abstract/Free Full Text]
-
Silverman PM, Roberts SC, Ducic I, et al. Assessment of a technology that permits individualized scan delays on helical hepatic CT: a technique to improve efficiency in use of contrast material. AJR Am J Roentgenol 1996; 167:79-84.[Abstract/Free Full Text]
-
Mitsuzaki K, Yamashita Y, Ogata I, Nishiharu T, Urata J, Takahashi M. Multiphase helical CT of the liver for detecting small hepatomas in patients with liver cirrhosis: contrast injection protocol and optimal timing. AJR Am J Roentgenol 1996; 167:753-757.[Abstract/Free Full Text]
-
Paulson EK, McDermott VG, Keogan MT, DeLong DM, Frederick MG, Nelson RC. Carcinoid metastases to the liver: role of triple phase helical CT. Radiology 1998; 206:143-150.[Abstract/Free Full Text]
-
Foley WD, Mallisee TA, Hohenwalter MD, Wilson CR, Quiroz FA, Taylor AJ. Multiphase hepatic CT with a multirow detector CT scanner. AJR Am J Roentgenol 2000; 175:679-685.[Abstract/Free Full Text]
-
Larson RE, Semelka RC, Bagley AS, Molina PL, Brown ED, Lee JKT. Hypervascular malignant liver lesions: comparison of various MR imaging pulse sequences and dynamic CT. Radiology 1994; 192:393-399.[Abstract/Free Full Text]
-
Kopka L, Rodenwaldt J, Fischer U, Mueller DW, Oestmann JW, Grabbe E. Dual phase helical CT of the liver: effects of bolus tracking and different volumes of contrast material. Radiology 1996; 201:321-326.[Abstract/Free Full Text]
-
Oliver JH, 3rd, Baron RL, Federle MP, Jones BC, Sheng R. Hypervascular liver metastases: do unenhanced and hepatic arterial phase CT images affect tumor detection?. Radiology 1997; 205:709-715.[Abstract/Free Full Text]
-
Baron RL, Oliver JH, 3rd, Dodd GD, 3rd, Nalesnik M, Holbert BL, Carr B. Hepatocellular carcinoma: evaluation with biphasic, contrast-enhanced, helical CT. Radiology 1996; 199:505-511.[Abstract/Free Full Text]
-
Weg N, Scheer MR, Gabor MP. Liver lesions: improved detection with dual detector array CT and routine 2.5 mm thin collimation. Radiology 1998; 209:417-426.[Abstract/Free Full Text]
-
Katyal S, Oliver JH, 3rd, Buck DG, Federle MP. Detection of vascular complications after liver transplantation: early experience in multislice CT angiography with volume rendering. AJR Am J Roentgenol 2000; 175:1735-1739.[Abstract/Free Full Text]
-
Kamel IR, Raptopoulos V, Pomfret EA, et al. Living adult right lobe liver transplantation: imaging before surgery with multidetector multiphase CT. AJR Am J Roentgenol 2000; 175:1141-1143.[Free Full Text]
-
Nelson RC, Johnson GA, Spielman AL, et al. Single breath-hold dynamic subtraction CT of the liver using multidetector helical technology (abstr). Radiology 1999; 213(P):125.
This article has been cited by other articles:

|
 |

|
 |
 
I. R. Kamel, L. P. Lawler, and E. K. Fishman
Comprehensive Analysis of Hypervascular Liver Lesions Using 16-MDCT and Advanced Image Processing
Am. J. Roentgenol.,
August 1, 2004;
183(2):
443 - 452.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. Catalano, F. Fraioli, A. Laghi, A. Napoli, F. Pediconi, M. Danti, P. Nardis, and R. Passariello
High-Resolution Multidetector CT in the Preoperative Evaluation of Patients with Renal Cell Carcinoma
Am. J. Roentgenol.,
May 1, 2003;
180(5):
1271 - 1277.
[Abstract]
[Full Text]
[PDF]
|
 |
|